Devices: Difference between revisions

Jump to navigation Jump to search
3,093 bytes added ,  9 May 2013
no edit summary
No edit summary
No edit summary
 
(26 intermediate revisions by 2 users not shown)
Line 1: Line 1:
==Pacemakers==
==Pacemakers==
[[Image: |right]]
[[Image:Paced2.gif|300px|right|thumb|Ventricular paced rhythm shows ventricular pacemaker spikes.]]
[[Image:Pacemaker_device.svg|right|thumb|300px|Schematic display of a pacemaker.]]


A pacemaker monitors the electrical impulses in the heart. When needed, it sends small electrical impulses to the heart muscle to maintain a normal heart rate.  
A pacemaker monitors the electrical impulses in the heart. When needed, it sends small electrical impulses to the heart muscle to maintain a normal heart rate.  


A pacemaker rhythm can easily be recognized on the ECG. It shows pacemaker spikes: vertical signals that represent the electrical activity of the pacemaker. Usually these spikes are more visible in unipolar than in bipolar pacing (see below).
A pacemaker rhythm can easily be recognized on the ECG. It shows pacemaker spikes: vertical signals that represent the electrical activity of the pacemaker. Usually these spikes are more visible in unipolar than in bipolar pacing (see image on the right).
 
[[Image:Paced2.gif ]]


==Patients==
==Patients==
Line 15: Line 14:


===Device measurements===
===Device measurements===
[[Image:RedTh.jpg|300px|right|thumb|Determination of stimulation threshold, after the red arrow the stimulus intensify and duration is not sufficient anymore to capture the heart.]]
{| class="wikitable" border="1" cellpadding="1" cellspacing="1" width="200px"
|-
!Ohm’s law:
|-
|U = I x R
''U = Voltage''
''I = Current''
''R = Impedance''
|}
====Stimulation threshold (in mV@ms):====  
====Stimulation threshold (in mV@ms):====  
The stimulation threshold is the minimum stimulus intensity (in mV) and duration (in ms) necessary to capture the heart.
The stimulation threshold is the minimum stimulus intensity (in mV) and duration (in ms) necessary to capture the heart.
Line 21: Line 36:
The degree that the pacing system “sees” or senses signals, controlled by the sensitivity setting which is graduated in mV.
The degree that the pacing system “sees” or senses signals, controlled by the sensitivity setting which is graduated in mV.


====Impedance (in ?):====  
====Impedance (in Ω):====  
The total pacing impedance is the sum of all resistance to the flow of the electrical impulses.  
The total pacing impedance is the sum of all resistance to the flow of the electrical impulses.  


Line 29: Line 44:
If these tests are within limits, the pacemaker leads are properly positioned. Thereafter, the pacemaker pulse generator is implanted subcutaneously or subpectorally under the left or right clavicle.
If these tests are within limits, the pacemaker leads are properly positioned. Thereafter, the pacemaker pulse generator is implanted subcutaneously or subpectorally under the left or right clavicle.


==Programming===
==Programming==
===Pacemaker codes===
===Pacemaker codes===
Pacemakers can be categorized according to the NASPE coding system that usually consists of 3-5 letters.
Pacemakers can be categorized according to the NASPE coding system that usually consists of 3-5 letters.


{| class="wikitable" border="1" cellpadding="1" cellspacing="1" width="80%"
{| class="wikitable" border="1" cellpadding="1" cellspacing="1" width="400px"
|-
|-
!I
!I
Line 85: Line 100:


===Explanations of table:===
===Explanations of table:===
====III====
====III:====
=====Triggered:=====  
{| border="0" cellpadding="1" cellspacing="1" width="100%"
|-
|bgcolor="#FAF8CC"|
=====Triggered:=====
|}
A sensed event triggers a pacemaker output pulse  
A sensed event triggers a pacemaker output pulse  
{| border="0" cellpadding="1" cellspacing="1" width="100%"
|-
|bgcolor="#FAF8CC"|
=====Inhibited:=====
=====Inhibited:=====
|}
Detection of physiological heart activitity will inhibit an electrical pacemaker impulse  
Detection of physiological heart activitity will inhibit an electrical pacemaker impulse  
{| border="0" cellpadding="1" cellspacing="1" width="100%"
|-
|bgcolor="#FAF8CC"|
=====Dual:=====
=====Dual:=====
|}
A pacemaker with dual response to sensing will inhibit a pacemaker output pulse if it senses an intrinsic event in that same chamber, but it will trigger a pacemaker output pulse in the ventricle if it senses an intrinsic event in the atrium (after a programmed atrioventricular interval)
A pacemaker with dual response to sensing will inhibit a pacemaker output pulse if it senses an intrinsic event in that same chamber, but it will trigger a pacemaker output pulse in the ventricle if it senses an intrinsic event in the atrium (after a programmed atrioventricular interval)
====IV:====
====IV:====
{| border="0" cellpadding="1" cellspacing="1" width="100%"
|-
|bgcolor="#FAF8CC"|
=====Rate modulation:=====
=====Rate modulation:=====
|}
In some patients, rate adaptive pacing is programmed “on” to ensure that when patients exercise increases, the pacemaker ensures that the heart rate increases to provide additional cardiac output. There are many ways to sense physiological exercise, including motion sensors and ventilation sensors.
In some patients, rate adaptive pacing is programmed “on” to ensure that when patients exercise increases, the pacemaker ensures that the heart rate increases to provide additional cardiac output. There are many ways to sense physiological exercise, including motion sensors and ventilation sensors.


===Commonly used pacemakers===
===Commonly used pacemakers===
[[Image:Ddd paced 12lead.jpg|300px|thumb|DDD paced rhythm]]
====AAI:====  
====AAI:====  
The atria are paced, when the intrinsic atrial rhythm falls below the pacemaker's threshold.  
The atria are paced, when the intrinsic atrial rhythm falls below the pacemaker's threshold.  
Line 109: Line 142:
Sensing and stimulation of the myocardium demands a closed electrical circuit. A pacemaker can sense and stimulate in a unipolar and bipolar fashion:
Sensing and stimulation of the myocardium demands a closed electrical circuit. A pacemaker can sense and stimulate in a unipolar and bipolar fashion:
====Unipolar:====  
====Unipolar:====  
[[Image:ECGT.jpg|thumb|right|700px|ECG with unipolar stimulation displaying large pacemaker spikes.            ECG with bipolar stimulation displaying small pacemaker spikes]]
Pacemaker circuit between 1 electrode at the distal end of lead and the pacemaker pulse generator (large circuit ± 40-60 cm).
Pacemaker circuit between 1 electrode at the distal end of lead and the pacemaker pulse generator (large circuit ± 40-60 cm).
*Advantage: large pacemaker spikes (easier interpretation pacemaker ECG)
*Advantage: large pacemaker spikes (easier interpretation pacemaker ECG)
*Disadvantage: extracardiac stimulation (pectoral muscle), sensing of extracardiac signals (such as ventricular depolarisations in the atrial sensing channel (far field R waves) or non-physiological noise)
*Disadvantage: extracardiac stimulation (pectoral muscle), sensing of extracardiac signals (such as ventricular depolarisations in the atrial sensing channel (far field R waves) or non-physiological noise)
====Bipolar:====  
====Bipolar:====  
Pacemaker circuit between 2 electrodes at the distal end of the lead (small circuit ± 10-15 mm)
Pacemaker circuit between 2 electrodes at the distal end of the lead (small circuit ± 10-15 mm)
Line 124: Line 160:
==Complications==
==Complications==
===Undersensing:===  
===Undersensing:===  
[[Image:Ventricular undersensing.svg]]
[[Image:Ventricular undersensing.svg|right|thumb|250px|Ventricular undersensing. Failure of appropriate ventriculation inhibition]]


An intrinsic depolarization that is present, is not sensed by the pacemaker. This can be due to:
An intrinsic depolarization that is present, is not sensed by the pacemaker. This can be due to:
Line 133: Line 169:


===Oversensing:===  
===Oversensing:===  
[[Image:Ventricular oversensing.svg]]
[[Image:Ventricular oversensing.svg|right|thumb|250px|Ventricular oversensing. Failure of appropriate ventricular firing]]


The pacemakers senses signals on the marker  
The pacemakers senses signals on the marker channel that do not correspond to the ECG pattern. This can be physiologic (e.g. ventricular pulse or myopotentials) or non-physiologic (e.g. lead fracture or if the lead is loose from the pacemaker pulse generator or outside interference such as TENS therapy or surgical diathermy).
channel that do not correspond to the ECG pattern. This can be  
physiologic (e.g. ventricular pulse or myopotentials) or  
non-physiologic (e.g. lead fracture or if the lead is loose from  
the pacemaker pulse generator or outside interference such as  
TENS therapy or surgical diathermy).


===Non-capture:===  
===Non-capture:===  
[[Image:AtrialCap.jpg|right|thumb|300px|Failure of atrial capture in a patient with atrial standstill, no P-waves are seen after the atrial stimuli.]]
The electrical impulse is not followed by the myocardium.
The electrical impulse is not followed by the myocardium.
[[Image:Ddd paced 12lead.jpg]]


This can be due to:
This can be due to:
Line 180: Line 211:


==Implantable cardioverter defibrillators (ICD)==
==Implantable cardioverter defibrillators (ICD)==
[[Image:ICD_device.svg|right|300px|thumb|Schematic display of an ICD]]
[[Image:Subcutaneous_ICD.svg|right|300px|thumb|A fully subcutaneous ICD (SQICD) has a can and lead that are placed subcutaneous but outside of the thorax. This type of ICD does not have te option to give anti-tachypacing or continuous pacemaker functionality, but can deliver a shock to cardiovert ventricular fibrillation.]]
An ICD is a device that monitors heart rhythms. If it senses dangerous rhythms, it delivers shocks or anti-tachypacing (ATP) therapy. Many ICDs record the heart's electrical patterns when there is an abnormal heartbeat.
An ICD is a device that monitors heart rhythms. If it senses dangerous rhythms, it delivers shocks or anti-tachypacing (ATP) therapy. Many ICDs record the heart's electrical patterns when there is an abnormal heartbeat.


Line 195: Line 230:
Patients do not meet the evidence based ICD implantation criteria if they have (1) a myocardial infarction within 40 days before ICD implantation; (2) newly diagnosed heart failure at the time of ICD implantation without prior therapy; (3) NYHA class IV symptoms of congestive heart failure.
Patients do not meet the evidence based ICD implantation criteria if they have (1) a myocardial infarction within 40 days before ICD implantation; (2) newly diagnosed heart failure at the time of ICD implantation without prior therapy; (3) NYHA class IV symptoms of congestive heart failure.


===Implantation===
A seperate chapter deals with a more complete list of [[ICD indications]]
 
==Implantation==
[[Image:Xthorax.jpg|right|300px|thumb|X-thorax of a patient with an ICD (in posteroanterior and lateral view)]]
 
ICDs are implanted under local anaesthesia in a sterile operating room or catheterization laboratory. Transvenous ICD implantation starts, like pacemaker implantation, with inserting the ICD leads in the left or right cephalic vein. Fluoroscopy is used to ensure the appropriate location in the atrium or ventricle. Leads are often positioned in the auricle of the right atrium and/or the right ventricular apex.  
ICDs are implanted under local anaesthesia in a sterile operating room or catheterization laboratory. Transvenous ICD implantation starts, like pacemaker implantation, with inserting the ICD leads in the left or right cephalic vein. Fluoroscopy is used to ensure the appropriate location in the atrium or ventricle. Leads are often positioned in the auricle of the right atrium and/or the right ventricular apex.  


Line 220: Line 259:
===Arrhythmia detection zones===
===Arrhythmia detection zones===
The ICD diagnoses rhythm disorders by counting intervals on the intracardiac electrogram. This is a rate-based detection scheme that can be adjusted to meet the individual patient’s needs by programming. The ICD counts the current interval as one value and then average of the current interval and the preceding intervals. If these intervals fall into the same category, the event is binned in that category. If both events are tachycardia of fibrillation, but not in the same category, the interval is binned in the higher category.  
The ICD diagnoses rhythm disorders by counting intervals on the intracardiac electrogram. This is a rate-based detection scheme that can be adjusted to meet the individual patient’s needs by programming. The ICD counts the current interval as one value and then average of the current interval and the preceding intervals. If these intervals fall into the same category, the event is binned in that category. If both events are tachycardia of fibrillation, but not in the same category, the interval is binned in the higher category.  
{| class="wikitable" border="1" cellpadding="1" cellspacing="1" width="600px"
|-
!SVT discriminators:
|-
|
#Waveform morphology (broad vs. small complex or comparison of morphology with template of normal QRS wave)
#Onset of arrhytmia (sudden vs. slowly)
#Stability of arrhythmia (regular vs. irregular)
#Relationship between P- and R-waves (atrial lead required)
|}


The arrhythmia detection zone is the category in which a predefined therapy will be given:
The arrhythmia detection zone is the category in which a predefined therapy will be given:
*'''Monitor zone (e.g. 160-180 bpm):''' All events in this zone will be recorded in the ICD and can be seen during follow up visits, however no therapy is given.
*'''Monitor zone (e.g. 160-180 bpm):''' All events in this zone will be recorded in the ICD and can be seen during follow up visits, however no therapy is given.
[[Image:ICD.jpg|right|thumb|300px|ICD strook met intracardiale electrocardiogram (EGM): VF met shock. De rode pijl geeft het moment van de shock aan.]]
*'''Fast VT zone (e.g. 180-240 bpm):''' All events in this zone will be recorded in the ICD and can be seen during follow up visits. Therapy is given or not if the arrhythmia satisfies several criteria, which are programmed as well, such as SVT/VT discriminators.
*'''Fast VT zone (e.g. 180-240 bpm):''' All events in this zone will be recorded in the ICD and can be seen during follow up visits. Therapy is given or not if the arrhythmia satisfies several criteria, which are programmed as well, such as SVT/VT discriminators.
*'''VF zone (e.g. >240 bpm):''' All events in this zone will be recorded in the ICD and can be seen during follow up visits and therapy is given immediately.
*'''VF zone (e.g. >240 bpm):''' All events in this zone will be recorded in the ICD and can be seen during follow up visits and therapy is given immediately.
Line 247: Line 298:


==Cardiac resynchronisation therapy (CRT)==
==Cardiac resynchronisation therapy (CRT)==
[[Image:CRT_device.svg|right|thumb|400px|Schematic display of an CRT device]]
[[Image:ChestXray.jpg|right|thumb|250px|Chest X-ray of a patient with an CRT device (in posteroanterior and lateral view)<br/>The red arrow is the right atrial lead<br/>The blue arrow is the right ventricular lead<br/>The green arrow is the coronary sinus lead]]
CRT-pacemaker (CRT-P) is a biventricular pacemaker with leads in both ventricles to ensure synchronized contraction. A CRT-defibrillator (CRT-D) is an ICD with biventricular pacing option.
CRT-pacemaker (CRT-P) is a biventricular pacemaker with leads in both ventricles to ensure synchronized contraction. A CRT-defibrillator (CRT-D) is an ICD with biventricular pacing option.
It appears that atrio-ventricular and intraventricular conduction delays further aggrevates left ventricular (LV) dysfunction in patients with underlying cardiomyopathies. Notably, left bundle branch block (LBBB) alters the sequence of LV contraction, causing wall segments to contract early or late. Dyssynchrony seems to represent a pathophysiological process that directly depresses ventricular function, causing LV remodelling and congestive heart failure and as a consequence causes a higher risk of morbidity and mortality. Atrio-biventricular pacing (CRT) for patients with symptomatic heart failure and intra- or interventricular conduction disturbances has proved beneficial.  
It appears that atrio-ventricular and intraventricular conduction delays further aggrevates left ventricular (LV) dysfunction in patients with underlying cardiomyopathies. Notably, left bundle branch block (LBBB) alters the sequence of LV contraction, causing wall segments to contract early or late. Dyssynchrony seems to represent a pathophysiological process that directly depresses ventricular function, causing LV remodelling and congestive heart failure and as a consequence causes a higher risk of morbidity and mortality. Atrio-biventricular pacing (CRT) for patients with symptomatic heart failure and intra- or interventricular conduction disturbances has proved beneficial.  
Line 259: Line 313:
*Right atrial lead
*Right atrial lead
*Right ventricular lead
*Right ventricular lead
*Left ventricular lead (positioned in the coronary sinus)
*Left ventricular lead. The left ventricular lead is usually positioned in the coronary sinus, alternatively it can be positioned epicardially on the left ventricle (by a surgical procedure) or intracardially in the left ventricle (through a transseptal puncture).
 
===Programming, follow up and complications===
===Programming, follow up and complications===
Programming should specifically aim at ensuring atrial-synchronous permanent biventricular pacing, by performing AV-interval optimization (echocardiography guided or using invasive haemodynamic measurments) and performing ventricular-ventricle (VV) interval optimization. Further programming, follow up and complications are similar to pacemakers and ICDs (see above).
Programming should specifically aim at ensuring atrial-synchronous permanent biventricular pacing, by performing AV-interval optimization (echocardiography guided or using invasive haemodynamic measurments) and performing ventricular-ventricle (VV) interval optimization. Further programming, follow up and complications are similar to pacemakers and ICDs (see above).

Navigation menu